The Efficacy of In-Phase and Quadrature Demodulation in Electronic Fetal Heart Rate Monitoring During Labor
文献摘要:
Objective::To investigate the efficacy of in-phase and quadrature (IQ) demodulation in electronic fetal heart rate monitoring (EFM) to reduce false reports of fetal heart rate (FHR) doubling or halving.Methods::This is a prospective cohort study. A total of 263 full-term pregnant women who delivered at Peking University Shenzhen Hospital between August 2019 and July 2020 were prospectively enrolled in the study. FHR monitoring began when the cervix was dilated to 2-3 cm and continued until delivery. Raw fetal Doppler audio signals and internal and external cardiotocography curves from internal electrode monitoring, EFM with conventional demodulation (external), and EFM with IQ demodulation (external) were acquired to compare FHR doubling and halving time. In cohort 1, FHR was compared between IQ demodulation and conventional demodulation. In cohort 2, FHR was compared between IQ demodulation, conventional demodulation, and internal FHR monitoring. Count data were statistically analyzed using the Chi-squared test, and measurement data were statistically analyzed using
t-test for correlation coefficients, and Bland-Altman analysis for concordance ranges.
Results::To compare IQ demodulation and conventional demodulation, 225 pregnant women were monitored for a total of 835,870 seconds. The beat-to-beat interval of FHRs in raw fetal Doppler audio signals was used as the reference. The results showed a doubling time of 3401 seconds (0.407%, 3401/835,870) and a halving time of 2918 seconds (0.349%, 2918/835,870) with conventional demodulation, compared to 241 seconds (0.029%, 241/835,870) and 589 seconds (0.070%, 589/835,870), respectively, with IQ demodulation. IQ demodulation reduced FHR doubling by approximately 93% (3160/3401) and FHR halving by approximately 80% (2329/2918) compared to conventional demodulation (
P < 0.01).
Conclusion::EFM with IQ demodulation significantly reduces false FHR doubling and halving, with an efficacy similar to that of internal FHR monitoring.
文献关键词:
Fetal monitoring;Heart rate, fetal;Doubling;Halving;IQ demodulation
中图分类号:
作者姓名:
Liang Yiheng;Liu Ping;Yan Shaomei;Li Yun;Chen Duijin;Fan Shangrong
作者机构:
Department of Obstetrics and Gynecology, Peking University Shenzhen Hospital, Shenzhen, Guangdong 518036, China;Shenzhen Key Laboratory on Technology for Early Diagnosis of Major Gynecological Diseases, Shenzhen, Guangdong 518036, China;Department of Obstetrics and Gynecology, The Third Affiliated Hospital of Guangzhou Medical University, Guangzhou, Guangdong 510150, China
文献出处:
引用格式:
[1]Liang Yiheng;Liu Ping;Yan Shaomei;Li Yun;Chen Duijin;Fan Shangrong-.The Efficacy of In-Phase and Quadrature Demodulation in Electronic Fetal Heart Rate Monitoring During Labor)[J].母胎医学杂志(英文),2022(02):113-120
A类:
halving,cardiotocography,FHRs,Halving
B类:
Efficacy,In,Phase,Quadrature,Demodulation,Electronic,Fetal,Heart,Rate,Monitoring,During,Labor,Objective,To,investigate,efficacy,phase,quadrature,IQ,demodulation,electronic,fetal,heart,rate,monitoring,EFM,false,reports,doubling,Methods,This,cohort,study,total,full,term,pregnant,women,who,delivered,Peking,University,Shenzhen,Hospital,between,August,July,were,prospectively,enrolled,began,when,cervix,was,dilated,continued,until,delivery,Raw,Doppler,audio,signals,internal,external,curves,from,electrode,conventional,acquired,compared,Count,data,statistically,analyzed,using,Chi,squared,test,measurement,correlation,coefficients,Bland,Altman,analysis,concordance,ranges,Results,monitored,seconds,beat,interval,raw,used,reference,results,showed,respectively,reduced,by,approximately,Conclusion,significantly,reduces,similar,that,Doubling
AB值:
0.403753
相似文献
机标中图分类号,由域田数据科技根据网络公开资料自动分析生成,仅供学习研究参考。