典型文献
A calibration-free model for laser-induced breakdown spectroscopy using non-gated detectors
文献摘要:
Calibration-free(CF)laser-induced breakdown spectroscopy(LIBS)is normally only applicable for gated detectors due to its dependence on the assumption of a steady-state plasma.However,most currently available LIBS systems are equipped with non-gated detectors such as charge-coupled device(CCD),which degrades the accuracy of CF method.In this paper,the reason for the less satisfactory quantification performance of CF for LIBS with non-gated detectors was clarified and a time-integration cali-bration-free(TICF)model was proposed for applications with non-gated detectors.It was based on an assumed temporal profile of plasma proper-ties,including temperature and electron density,obtained from another pre-experiment.The line intensity at different time during the signal collection time window was estimated with self-absorption correction according to the temporal profile of the plasma properties.The proposed model was validated on titanium alloys and compared with traditional CF.The accuracy of elemental concentration measurement was improved significantly:the average relative error of aluminum and vanadium decreased from 6.07%and 22.34%to 2.01%and 1.92%,respectively.The quantification results showed that TICF method was able to extend the applicability of CF to LIBS with non-gated detectors.
文献关键词:
中图分类号:
作者姓名:
Zongyu Hou;Weilun Gu;Tianqi Li;Zhe Wang;Liang Li;Xiang Yu;Yecai Zhang;Zijun Liu
作者机构:
State Key Lab of Power System,Department of Energy and Power Engineering,International Joint Laboratory on Low Carbon Clean Energy Innovation,Tsinghua University,Beijing 100084,China;Shanxi Research Institute for Clean Energy,Tsinghua University,Taiyuan 030032,China;Beijing Research Institute of Chemical Engineering and Metallurgy,Beijing 101149,China;China National Uranium Corporation,Beijing 100013,China;Guoneng Shenwan Energy Co.,Ltd,China;Jinneng Holding Tashan Power Generation Co.,Ltd,China
文献出处:
引用格式:
[1]Zongyu Hou;Weilun Gu;Tianqi Li;Zhe Wang;Liang Li;Xiang Yu;Yecai Zhang;Zijun Liu-.A calibration-free model for laser-induced breakdown spectroscopy using non-gated detectors)[J].物理学前沿,2022(06):167-175
A类:
TICF
B类:
calibration,free,model,laser,induced,breakdown,spectroscopy,using,gated,detectors,Calibration,LIBS,normally,only,applicable,due,its,dependence,assumption,steady,state,plasma,However,most,currently,available,systems,equipped,such,charge,coupled,device,CCD,which,degrades,accuracy,method,In,this,paper,reason,less,satisfactory,quantification,performance,was,clarified,integration,proposed,applications,It,assumed,temporal,profile,including,temperature,electron,density,obtained,from,another,pre,experiment,line,intensity,different,during,signal,collection,window,estimated,self,absorption,correction,according,properties,validated,titanium,alloys,compared,traditional,elemental,concentration,measurement,improved,significantly,average,relative,error,aluminum,vanadium,decreased,respectively,results,showed,that,extend,applicability
AB值:
0.533453
相似文献
机标中图分类号,由域田数据科技根据网络公开资料自动分析生成,仅供学习研究参考。